OPERATING
INSTRUCTIONS FOR CLD SERIES
LIQUID DISTRIBUTION SYSTEMS
Introduction
The container
consists of an inner stainless steel cylinder securely supported
in an outer jacket shell. The space between the inner and
outer vessels contains a highly efficient insulation material
and is evacuated. The container has four (4) valves: A liquid
valve (V-1), a full trycock valve (V-2), a pressure building
valve (V-3), and a vent valve (V-4). A pressure-building regulator
(PVC-1) is located in series with the pressure-building valve.
Safety
Devices
A safety
relief valve and a rupture disc on the vent manifold protect
the inner liquid reservoir.
A combination
evacuation valve and relief device is provided to service
the vacuum space. This protects the container in the event
of a leak in the inner reservoir. If this device vents, contact
Cryofab; do not attempt to use the container or re-evacuate
the insulation space.
Gauges
A pressure
gauge is provided indicating inner vessel pressure. Aliquid
level gauge is provided to indicate approximate container
contents.
Handling
These
containers are designed for use in the upright position and
must never be laid on their side. If a container must be lifted
use a forklift or similar device beneath the base, or hoist
by means of the lifting lugs on the top head. Do not attempt
to lift by means of slings around the shell.
Pressure-Building
System
This container
is equipped with an integral pressure-building system to aid
in liquid withdrawal as long as a sufficient liquid supply
is in the container and the pressure-building valve open.
If a different pressure setting is required, the pressure-building
regulator can be adjusted. To lower the setting, turn the
adjusting screw counterclockwise. To increase the setting,
turn clockwise.
Filling
These
containers can be filled from a pressurized liquid source
using the following procedure. Attach the extension to the
full trycock valve (V-2, located top center) and orient it
so that the discharge is directed away from personnel and
out of van compartment. Connect the liquid source to the fill/withdrawal
valve, (V-1, labeled liquid) using a suitable transfer line.
Open the full trycock valve and the fill/withdrawal valve.
To begin the transfer open the liquid source valve.
WARNING
A cold
stream of gas and or liquid will exit from the full trycock
extension. Keep clear of exiting stream.
When the
container is filled to maximum capacity, liquid will start
to exit from the full trycock; at that point, shut off the
liquid source valve. (Consult DOT regulations or your local
liquid supplier for the maximum allowable liquid load consistent
with your working pressure). Disconnect the transfer line
and the full trycock extension (if necessary). Shut off the
fill/withdrawal valve and the full trycock valve.
Withdrawal
Transferring
liquid from the container is accomplished by the following
procedure: Be sure the vent valve (V-4, located on the vent
manifold) is closed and check the pressure gauge to see that
the vessel pressure is adequate for the intended application.
If additional pressurization is required, open the pressure-building
valve (V-3) to operate the integral pressure-building coil
(PBC-1). With the pressure building system functioning, the
preset pressure will be maintained throughout the withdrawal,
as long as sufficient liquid supply is present.
Attach
a suitable transfer line to the fill/withdrawal valve (V-1).
Open the fill/withdrawal valve as far as necessary to obtain
the desired flow rate. When the transfer is complete, close
the fill/withdrawal valve and disconnect the transfer line.
The pressure-building valve may be left open, if desired.
This may, to a limited extent, determine the operating pressure
of the vessel until the liquid is saturated.
WARNING
Do not
operate this vessel in a closed compartment or confined area
unless the vent stack is ducted to the outside. Do not open
the fill/withdrawal or the full trycock valve if in a confined
area.
Changing
Service
CLD series
are suitable for carrying liquid oxygen, nitrogen or argon,
however, changing from inert service to oxygen service is
not recommended.
CAUTION
Before
operating this vessel, please read thoroughly, "PRECAUTIONS
FOR THE SAFE HANDLING AND STORAGE OF LIQUEFIED GASES",
included with these instructions, for additional precautions
and good practice regarding cryogenics.
|